PRODUCT

Karl Fischer Moisture Titrator (Coulometric titration)
MKC-710M

Large color TFT-LCD with a touch panel (8.4-inch) enables easy key entry
Up to 4 titrators, AT-710B potentiometric titrators, MKV-710B Volumetric or MKH-710/2nd Hybrid or additional MKC-710B Coulometric Karl Fischer moisture titrators can be connected and simultaneous measurement is possible
Reduction of power consumption by 30% compared with the conventional product

Inquiry

Features

・Fast measurements

Our proprietary technology achieves electrolytic speeds up to 2.6mg H2O/min.
This shortens the time required for pre-titrations and sample measurements considerably.

・Replaceable diaphragm

Easy maintenance when measuring samples which tend to contaminate the diaphragm as eg. oils: Thanks to a unique mechanism, the ceramic diaphragm of the optional titration cell unit (12-03635-01) can be replaced.

・Result output as PDF files

Paper saving and environmentally friendly – results no longer need to be printed.
Measurement results are converted to PDF and can be stored in a USB flash drive.

  • The MKC-710M contains functions to support compliance with the following:

・European Pharmacopoeia Chapter 2.5.32. WATER: MICRO DETERMINATION: VERIFICATION OF ACCURACY
・Japanese Pharmacopoeia Eighteenth Edition Chapter 2.48: Water determination (Karl Fischer Method):
 2.4 Suitability of determination

  • Up to 4 simultaneous measurements (Simultaneous titrations)
  • Can be connected to AT-710B potentiometric titrators, MKV-710B Volumetric Karl Fischer moisture titrators or MKH-710/2nd Hybrid Karl Fischer moisture titrators
  • The main control unit and a titrator are separately located by wireless (Bluetooth(R))
  • Large color TFT-LCD with a touch panel (8.4-inch)
  • Two different user levels and permissions
  • With a unique mechanism , the ceramic diaphragm of the optional titration cell unit can be replaced
  • Measurement results are converted to PDF and can be stored in a USB flash drive
  • By reviewing the conventional electrolytic control method , reduction of power consumption by 30%
  • When using ADP-611 , a patented scan mode automatically determines the optimal evaporation temperature. (Japanese Patent No. 4247093)
  • Manual solvent change unit is included

Specifications

Type Karl Fischer Moisture Titrator
Model MKC-710M
Standard configuration MCU-710M+MKC-710+IDP-100+Manual Solvent Change unit
Measuring method Karl Fischer Coulometric titration
Measuring range Water content : 1ug to 300mg (depends on reagent)
Bromine index : 8ug to 300mg
Measurement cell 2-Component or 1-Component
Precision Relative standard deviation : less than 0.3% (n=10)
*Per KEM standard measurement conditions and standard liquids
Display resolution 0.1ug
Control method Constant current pulse time control
Endpoint detection Alternate current polarization method with a twin platinum electrode
EP sense method Selective drift stability or limit measurement time
Required solvent 2-Component cell : Anolyte 100mL , Catholyte 5mL
1-Component cell : Anolyte 150mL
Method 120
Key operation Touch panel
Displays 1) 8.4-inch color LCD 800 x 600 dots
2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish / German / French
3) Simultaneous 4-channel display (Can also display Automatic Potentiometric Titrator simultaneously)
Calculation Concentration of water content , statistics data processing (mean , SD and RSD) and automatic averaging of blank value and factor value
Data storage 500 samples
GLP conformance Registration of operator / User group administration
Check performance with standard substance :
Reminder of scheduled check date / Record of check results
Reagent life control : Reminder of expiration/Reminder reagent replacement date
Management of conduction time : Display of operating time
External I/O RS-232C port x4 :
for Dot matrix printer , Electronic balance , Data Capture Software (SOFT-CAP) , Evaporator , Multiple sample changer

USB ×1 :
for USB flash drive , Thermal printer , A4 printer , Keyboard , Barcode reader , Foot switch , USB HUB

LAN ×1 :
for Personal computer (PC)
Extensibility Measuring instrument :
Automatic Potentiometric Titrator (AT-710) , Karl Fischer Moisture Titrator (MKV-710/MKC-710/MKH-710)
Three of these instruments can be added

Evaporator ADP-611

Multiple sample changer : CHK-501
Ambient condition 1) Temperature : 5 to 35°C
2) Humidity : 85%RH or below (no condensation)
Power source AC100 to 240V +/- 10% 50/60 Hz
Power consumption Main unit : Approx. 30W Printer : Approx. 7W
Dimensions Touch-on panel controller : 225(W) x 190(D) x 42(H) mm
Titration unit : 141(W) x 292(D) x 244(H) mm
Stirrer : 107(W) x 206(D) x 340(H) mm (not incl. Solvent Change unit)
Solvent Change Unit : 240(W) x 140(D) x 405(H) mm (not incl. tubing)
Printer : 106(W) x 180(D) x 88(H) mm
Weight Touch-on panel controller : Approx.1.5kg Titration unit : Approx.3.0kg Stirrer : Approx.2.0kg Solvent Change Unit : Approx.0.6kg Printer : Approx.0.4kg
Conformity standard CE marking EMC : EN61326-1 LVD : EN61010-1

Standards

ASTM D1123 Standard Test Methods for Water in Engine Coolant Concentrate by the Karl Fischer Reagent Method


ASTM D1533 Standard Test Method for Water in Insulating Liquids by Coulometric Karl Fischer Titration


ASTM D3401 Standard Test Methods for Water in Halogenated Organic Solvents and Their Admixtures


ASTM D4672 Standard Test Method for Polyurethane Raw Materials: Determination of Water Content of Polyols


ASTM D4928 Standard Test Method for Water in Crude Oils by Coulometric Karl Fischer Titration


ASTM D5460 Standard Test Method for Rubber Compounding Materials—Water in Rubber Additives


ASTM D6304 Standard Test Method for Determination of Water in Petroleum Products, Lubricating Oils, and Additives by Coulometric Karl Fischer Titration


ASTM D6869 Standard Test Method for Coulometric and Volumetric Determination of Moisture in Plastics Using the Karl Fischer Reaction (the Reaction of Iodine with Water)


ASTM E1064 Standard Test Method for Water in Organic Liquids by Coulometric Karl Fischer Titration


European Pharmacopoeia / 2.5.32. WATER: MICRO DETERMINATION


ISO 10101-3 Natural gas -- Determination of water by the Karl Fischer method -- Part 3: Coulometric procedure


ISO 12937 Petroleum products -- Determination of water -- Coulometric Karl Fischer titration method


ISO 14897 Plastics -- Polyols for use in the production of polyurethane -- Determination of water content


ISO 15512 Plastics -- Determination of water content


ISO 760 Determination of water -- Karl Fischer method (General method)


JIS C2101 Testing methods of electrical insulating oils


JIS K0068 Test methods for water content of chemical products


JIS K0113 General rules for methods of potentiometric, amperometric, coulometric, and Karl Fischer titrations


JIS K1557-2 Plastics -- Polyols for use in the production of polyurethane -- Part 2: Determination of water content


JIS K2275-3 Crude petroleum and petroleum products -- Determination of water -- Part 3: Coulometric Karl Fischer titration method


JIS M8211 Iron ores -- Method for determination of combined water content


JIS M8250 Iron ores -- Determination of hygroscopic moisture in analytical samples -- Gravimetric, Karl Fischer titration and mass-loss methods


United States Pharmacopeia–National Formulary / 38 <921> WATER: DETERMINATION


Software

Option

Catalog

■General-Catalog.pdf

■Karl Fischer Moisture Titrator MKV-710 MKC-710.pdf

■KEM-SET-for-Petrochemical.pdf

■KEM-SET-for-Palm-Oil.pdf

■KEM-SET-for-Food.pdf

■KEM-SET-for-Cosmetics.pdf

■KEM-SET-for-Beverage.pdf